Computation settings

Syntax

WU p1,p2,p3,p4<terminator>
p1 Setting type (MATH)
p2 Display on error

+OVER Positive overflow
-OVER Negative overflow

p3 Data when the SUM or AVE value overflows

ERROR Sets the computed result to the

value of the computation error

SKIP

Discards the data that overflowed
and continues the computation

LIMIT Processes the data as follows:

• For measurement channels

that do not have linear scaling
specified, the FX sets the data
to the upper or lower limit of the
measuring range.

• For measurement channels that

have linear scaling specified,
the FX sets the data to the
specified scan upper or lower
limit.

• For computation channels,

the FX sets the data to the
specified span upper or lower
limit.

p4 Data when the MAX, MIN, or P-P value

overflows
OVER

Uses the data that overflowed to
perform computations

SKIP

Discards the data that overflowed
and continues the computation

Description On models with the /M1, /PM1, or /PWR1 math

option, you can set p1 to MATH.

Report settings

Syntax

WU p1,p2,p3,p4,p5,p6<terminator>
p1 Setting type (REPORT)
p2 Report computation type 1

MAX

Maximum value

MIN

Minimum value

AVE

Average value

SUM

Integrated value

INST

Instantaneous value

p3 Report computation type 2

OFF

Disables report computation

MAX

Maximum value

MIN

Minimum value

AVE

Average value

SUM

Integrated value

INST

Instantaneous value

p4 Report computation type 3

The same as p3

p5 Report computation type 4

The same as p3

p6 How to create the “hourly+daily,”

“daily+weekly,” and “daily+monthly” files
COMBINE Saves reports to one file
SEPARATE Saves reports to separate files

Description • On models with the /M1, /PM1, or /PWR1

math option, you can set p1 to REPORT.

• For parameters p2 to p5, you cannot specify

the same computation type. The exception
is OFF, which can be specified for multiple
parameters. However, you cannot set all these
parameters to OFF.

Service ports

Syntax

WU p1,p2,p3,p4,p5<terminator>
p1 Setting type (SERVICEPORT)
p2 FTP service port (1 to 65535)
p3 Web service port (1 to 65535)
p4 SNTP service port (1 to 65535)
p5 Modbus service port (1 to 65535)

Description On models that have the Ethernet interface (/C7

option), you can set p1 to SERVICEPORT.

Decimal point type

Syntax

WU p1,p2<terminator>
p1 Setting type (DECIMALPOINT)
p2 Decimal point type (POINT, COMMA)

POINT A period is used for the decimal

point.

COMMA A comma is used for the decimal

point.