Note

The recalculation is done automatically when any change is made to any of

the :CALCulate[1|2] commands when :CALC:IMM:AUTO ON.

:CALCulate[1]:IMMediate?

Queries the results of the new calculation.

Note

As the commands INIT:CONT OFF and CALCulate[1]:IMMediate:AUTO OFF are inputted,

the command :CALCulate[1]:IMMediate? is just valid.

:CALCulate[1]:IMMediate:AUTO <Boolean>

Sets if the re-calculation will automatically be done whenever there’s any change made

to any of the :CALCulate[1|2] commands.

When it is set to OFF, the :CALCulate[1|2] command produces new results when the

command is executed, even when new data is not acquired. And this allows the user to

have the new results on the same current data right after new configuration changes.

Return

An ASCII-encoded byte. 0 means OFF, and 1 means ON.

Notes

Affects all the recalculation command settings.

This will be set to OFF after the counter has been turned off or a remote interface

reset.

:CALCulate[1]:MATH[:EXPRession]:CATalog?

Queries the defined equation name.

Return

Returns the string “SCALE_OFFSET”.

:CALCulate[1]:MATH[:EXPRession][:DEFine]?

Queries the equation used in the calculation.

Return