BNC 1105 Universal Counter User Manual

Page 49

Advertising
background image

49

49

49

49

To set upper or lower limit
1. Press LIMIT MODES button.
2. Press the softkey under UPPR to select upper limit or under LOWR to

select lower limit. Use the numeric keypad or the arrow keys to enter
a desired limit value (refer to Ch.4.2.1).
Upper Range:

-9.999,999,000,0E+12 to -1.000,000,000,0E-13

0.000,000,000,0

+1.000,000,000,0E-13 to +9.999,999,000,0E+12

Value Resolution: 11 digits

Lower Range:

-9.999,999,000,0E+12 to -1.000,000,000,0E-13

0.000,000,000,0

+1.000,000,000,0E-13 to +9.999,999,000,0E+12

Value Resolution: 11 digits


3. Press the softkey under DONE to finish setting the limits.

Remote Interface Command
Use the following commands to set upper/lower limits:

:CALCulate2:LIMit:UPPer[:DATA] <numeric value> [HZ|S|DEG]

:CALCulate2:LIMit:LOWer[:DATA] <numeric value> [HZ|S|DEG]

To use limit testing
1. Press LIMIT MODES button.
2. Press the softkey under ONFAIL to toggle between GO-ON and STOP.

Select GO-ON so that the Counter keeps making measurements when
the measurement exceeds the limits, or select STOP so that the
counter stops measuring when the measurement exceeds the limits.

3. Press the softkey under DONE to finish the setting.

Remote Interface Command
Use the following commands to activate the limit testing and to set
ONFAIL mode:

:CALCulate2:LIMit:STATe OFF|ON

:INITiate:AUTO OFF

:INITiate:AUTO ON