6 statistics operations – BNC 1105 Universal Counter User Manual

Page 50

Advertising
background image

50

50

50

50

To display graph for limit testing result
The counter display Limit Testing results in a graph where a circle within
two vertical bars represents a measurement within user-defined limits,
and the setting information will be showed. Users Show Status > Limit.
When users finish the Limit Testing settings and press the softkey under
DONE, the graph display mode will be presented at the bottom of the
display by pressing the softkey “Show Status”, and toggle between
options until “Limit” appears.

Remote Interface Command
Use the following command to select display mode for Limit Testing
result:

:CALCulate2:LIMit:DISPlay GRAPh|

NUMBer

To specify OUT*
Please refer to the instruction in Chapter 5.
1. Press Limit Modes button.
2. Press the softkey under OUT* to toggle between Off, USB or DB9
(1105-opt03)

Remote Interface Command

:SYSTem:COMMunicate:SERial:CONTrol:DTR OFF|LIM

:SYSTem:COMMunicate:SERial:SOURce USB|DB9


4.6 Statistics Operations

The MODEL 1105 may be activated to collect statistics on measurements
within specified limits or on other measurement except Peak Volts and
Totalize. Press the STATS button to enter the menu for setting up
statistics operations:

To enable/disable statistics operations: press the STATS in the
MATH area on the front panel to turn on or off the statistics
operation.

To select which statistic to display: press the softkey under
SHOW to toggle between Stdev, Max, Min, Mean, Delta, Meas and

Advertising