A USB or DB9 message will be returned.

Note

As the command :SYSTem:COMMunicate:SERial:SOURce USB is executed,

users

can’t do the remote control via USB excluding LAN or GPIB.

::::SYSTem:COMMunicate:SERial:CONTrol:DTR OFF|

|LIMit

Sets or queries the hardware pacing scheme. The parameter LIMit can be set by the

command :SYSTem:COMMunicate:SERial:SOURce USB|DB9. As the command is set to

OFF, users can’t use the Output function of the Limit PASS/FALL.

Return

A sequence of ASCII encoded bytes: OFF or LIM.

Notes

This value is stored at non-volatile memory, so it does not change after the counter

has been turned off or after a remote interface reset.

The start and stop thresholds are not user configurable.

:SYSTem:ERRor?

Queries the first error in the error queue and removes it from the error queue. Refer to

Chapter 6 for the error numbers and messages.

Return

The error number and the error message are returned. The error number is an

integer between -32767 and 32767. A “0” is returned if there is no error in the error

queue.

Notes

The error queue will be emptied after the counter has been turned off or after the last

error in the queue has been read.

When there are more than 30 errors, which causes the error queue overflows, the

last error is replaced with error -350, “queue overflows”. The most recent error will

be discarded.

This command removes the displayed error from the error queue.

The *RST command does not change the error queue.