BNC 1105 Universal Counter User Manual

Page 71

Advertising
background image

71

71

71

71

The MEASure? Command
Although it does not offer much flexibility, using the command :MEASure?
is the simplest way to program the Counter for measurement. With the
command, the Counter makes measurement according to the selected
measurement function and sends the result to the output buffer.

:MEASure[:SCALar]:<function>? <parameters> [,<source_list>]

For more information about <function>, <parameters> and <source_list>, please

refer to the above table.

•

If users skip the parameters and the function, or use the keyword DEFault,

the default setting (parameters and function) will be used.

•

Using this command automatically turns off the statistics, scale and offset

and limit testing operations.

•

If the trigger level is not specified in <parameters>, then auto-triggering will

be enabled and the trigger levels will be automatically set in all functions but

Voltage Peaks and Totalize on the measurement channel

.

:MEASure

:MEASure[:SCALar]

[:VOLTage]:FREQuency? [<expected_value> [,<resolution>]]

[,(@1)|(@2)|(@3)]

[:VOLTage]:FREQuency:RATio? [<expected_value> [,<resolution>]]

[,(@1),(@2)|(@3)]|[,(@2)|(@3),(@1)]

[:VOLTage]:PERiod? [<expected_value> [,<resolution>]]

[,(@1)|(@2)|(@3)]

[:VOLTage]:RISE:TIME? [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:RTIMe? [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:FALL:TIME? [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:FTIMe? [<lower_reference> [,upper_reference]][,(@1)]

[:VOLTage]:PWIDth? [<reference>][,(@1)]

[:VOLTage]:NWIDth? [<reference>][,(@1)]

[:VOLTage]:INITerval? [(@1),(@2)]

[:VOLTage]:PTPeak? [(@1),(@2)]

[:VOLTage]:MAXimum? [(@1),(@2)]

[:VOLTage]:MINimum? [(@1),(@2)]

[:VOLTage]:CENTer? [(@1),(@2)]