The command :configure – BNC 1105 Universal Counter User Manual

Page 120

Advertising
background image

120

120

120

120

Notes

The security state is stored in non-volatile memory and is unaffected by

power-on, save/recall or *RST.

The security code is set to 1105 when the Counter is shipped from the

factory. If users forget the security code, they can use

the :CALibration:SECurity:CODE:RESet command to recover to the default

condition.

:CALibration:SECurity:CODE:RESet

To prevent missing the user-defined code, and access the calibration mode

quickly, users can execute the command to recover to the default condition.

The Command :CONFigure

The commands :CONFigure and :READ offer more flexibility than the
command :MEASure. It provides the configuration part of the
measurement, but does not perform measurement nor query the result.
When using the comand :CONFigure, the Scale and Offset operation,
statistics and limit-testing are all disabled. If an absolute trigger level is
not specified in the <parameters>, auto-trigger and auto-trigger level
will be enabled and set for all measurement function except Totalize and
Voltage Peaks. Refer to (The Command :MEASure?) later in this section
for the descriptions of the <parameters> and the <source list> for each
function.

※Note: When you use this command, several default settings are automatically
used. In order to switch to another function without changing the settings, use

[:SENS]:FUNC[:ON]

:CONFigure

[:SCALar]:<function> <parameters> [,<source_list>]

To perform the specified function without initiating the measurement, please use the

command :INIT;FETC? or READ? to make and query a measurement. Skipping the

<parameters> and the [,<source_list>] will recover to the default condition.

:CONFigure?

Queries the function configured by the last :CONFigure or :MEASure command.

Return