The command :configure – BNC 1105 Universal Counter User Manual
Page 120
120
120
120
120
Notes
The security state is stored in non-volatile memory and is unaffected by
power-on, save/recall or *RST.
The security code is set to 1105 when the Counter is shipped from the
factory. If users forget the security code, they can use
the :CALibration:SECurity:CODE:RESet command to recover to the default
condition.
:CALibration:SECurity:CODE:RESet
To prevent missing the user-defined code, and access the calibration mode
quickly, users can execute the command to recover to the default condition.
The Command :CONFigure
The commands :CONFigure and :READ offer more flexibility than the
command :MEASure. It provides the configuration part of the
measurement, but does not perform measurement nor query the result.
When using the comand :CONFigure, the Scale and Offset operation,
statistics and limit-testing are all disabled. If an absolute trigger level is
not specified in the <parameters>, auto-trigger and auto-trigger level
will be enabled and set for all measurement function except Totalize and
Voltage Peaks. Refer to (The Command :MEASure?) later in this section
for the descriptions of the <parameters> and the <source list> for each
function.
※Note: When you use this command, several default settings are automatically
used. In order to switch to another function without changing the settings, use
[:SENS]:FUNC[:ON]
:CONFigure
[:SCALar]:<function> <parameters> [,<source_list>]
To perform the specified function without initiating the measurement, please use the
command :INIT;FETC? or READ? to make and query a measurement. Skipping the
<parameters> and the [,<source_list>] will recover to the default condition.
:CONFigure?
Queries the function configured by the last :CONFigure or :MEASure command.
Return