Starts an internal self-test and returns a response indicating if there is any error

found in the self-test.

Return

Numeric data transferred as ASCII bytes in <NR1> format.

A response value of zero indicates a complete self-test without any errors

and a response value of non-zero value indicates either a complete test with

errors or an unfinished self-test.

If the self-test fails, error -330 will be generated.

Note

The self-test includes the followings:

CPU, ROM, RAM, EEPROM, QSPI, FPGA, Front End, Measurement hardware

and Interpolator hardware.

*WAI

This command prevents the instrument from executing any further commands

or queries until the measurement cycle transitions from measuring to idle. The

only way to cancel this “holdoff” is by device memory clearance or power-on.

Note

This command is not affected by *RST and *CLS commands.

D.4 IEEE-488 Compliance Information

This section encloses a list of commands that are device-specific to the MODEL

1105. These commands are designed with the SCPI format in mind and they

follow

all the syntax rules of the standard.

:CALCulate[1]

:DATA?

:FEED “[:]SENSe[1]”

:IMMediate

:AUTO <Boolean>