BNC 1105 Universal Counter User Manual

Page 96

Advertising
background image

96

96

96

96

+25 dBm DC ±12V

External Arm Input Specifications

Signal Input Range

LVTTL and TTL compatible

Timing Restrictions

Pulse Width

> 50 ns

Transition Time

< 250 ns

Start-to-Stop Time

> 50 ns

Damage Level

12 Vrms

External Arm Input Characteristics

Impedance

1 kΩ

Input Capacitance

17 pF

Start Slope

Positive or Negative

Stop Slope

Positive or Negative

Notes

1. External Arm is available for all measurements except Peak Volts.
2. External Arm is referred to as External Gate for some measurements.

Internal Time Base Stability

Standard

(0° to 50°C)

High Stability Oven

(1105-opt01)

Temperature Stability

(referenced to 25°C)

± 1 x 10E-6

± 5 x 10E-9

Aging Rate

Per Day

Per Month

Per Year

± 0.2 x 10E-6 (Typical)

± 1 x 10E-6 (Typical)

± 8 x 10E-10

± 8 x 10E-8

Turn-on stability vs. time (30

min.)

± 2.0 x 10E-8

(refer to 24 hours)

Calibration

Electronic

Electronic

External Time Base Input Specifications

Voltage Range

200 mVrms to 10 Vrms

Damage Level

12 Vrms

External Time Base Input Characteristics

Threshold

0 V

Impedance

1 kΩ

Input Capacitance

25 pF

Input Frequency

10 MHz

Internal vs. External Time

Base Selection

Manual

Select Internal or External

Automatic

Internal used when External not present

(default)

Time Base Output Specifications

Output Frequency

10 MHz

Voltage

570 mVpp (0 dBm), typical

Impedance

50 Ω (typical), AC coupled

Measurement Specifications

Frequency, Period

1 mHz to 400 MHz (2.5 ns to 1000 s)

Advertising