BNC 1105 Universal Counter User Manual
Page 133
133
133
133
133
<source list>
Default
(@1)
Range
(@1)|(@2)
:MEASure[:SCALar][:VOLTage]:MINimum? [(@1)|(@2)]
Measures voltage minimum.
<source list>
Default
(@1)
Range
(@1)|(@2)
:MEASure[:SCALar][:VOLTage]:NWIDth? [<reference>][,(@1)]
Sets the measurement arming to “auto” and measures negative pulse width.
The <reference> specifies the point on the pulse where the negative pulse width is determined.
This point can be specified in percentage or in voltage. The default unit is percent.
The <reference> parameter is used to configure the trigger setting. When it is specified
in percentage or default, auto-trigger is enabled and trigger level is the specified value.
When it is specified in voltage, auto-trigger is disabled and the trigger level is the
specified value.
<reference>
default
50 pct
Percent range
0 to 100 pct
Percent resolution
10%
Voltage range
-5.125V to 5.125V for ×1 attenuation
-51.25V to 51,25V for ×10 attenuation
Voltage resolution
0.0025V for ×1 attenuation
0.025V for ×10 attenuation
<source list>
default
(@1)
range
(@1)
:MEASure[:SCALar][:VOLTage]:PWIDth? [<reference>][,(@1)]
Sets the measurement arming to “auto” and measures positive pulse width.
The <reference> specifies the point on the pulse where the positive pulse width is determined.
This point can be specified in percentage or in voltage. The default unit is percent.
The <reference> parameter is used to configure the trigger setting. When it is specified
in percentage or default, auto-trigger is enabled and trigger level is the specified value.
When it is specified in voltage, auto-trigger is disabled and the trigger level is the
specified value.
<reference>
default
50 pct
Percent range
0 to 100 pct
Percent resolution
10%