BNC 1105 Universal Counter User Manual

Page 149

Advertising
background image

149

149

149

149

negative edge.

Notes

This will be set to Positive after the counter has been turned off or a remote interface

reset.

This command only applies when [:SENSe]:TINT:ARM:ESTOP:LAY2:SOUR EXT is

used.

[:SENSe]:TINTerval:ARM:ESTOP:LAYer2:SLOPe IMMediate

︱EXTernal

Sets or queries the stopping point of the arm for Time interval measurements.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means

External.

Note

This will be set to Immediate after the counter has been turned off or a remote

interface reset.

[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:ECOunt <numeric value>

Sets or queries the number of events used to delay the stop arm for Time Interval

measurements. The range is from 1 to 99,999,999. The resolution is 1.

Return

Numeric data transferred as ASCII bytes in <NR1> format.

Notes

This will be set back to 1 after the counter has been turned off or a remote interface

reset.

This command only applies when [:SENSe]:TINT:ARM:ESTOP[:LAYer[1]]:SOUR

Internal2 is used.

[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:SOURce IMMediate

︱ TIMer ︱

INTernal2

Sets or queries the stopping point of the arm for Time Interval measurements.

Return