6.5 How to do the Gain Calibration via CH1

The Offset Calibration via CH1 is for checking DC +5V. Please do the
following procedures to finish the calibration.

1. Press UTILITY > CAL > Enter Security Code > Cal Diag > Gain

Inp1.

2. Use the BNC T-connector with 2 BNC cords to link an Agilent

33220A and a DMM (multimeter) as shown in the Figure 5-1.

3. Under the Agilent 33220A operation, the procedures should be

Utility > DC ON > Output Setup > Hi-Z.

4. Adjust the voltage on the Agilent 33220A, and let the DMM

displays +5.000V ±0.001V.

5. Remove the input source, from the DMM, on the counter.

6.

Press the softkey under RunCal to do the calibration. Then, the
counter’s display will show “Calibrating…”. The finished message
will disappear from the display as calibration is done. On the
contrary, the NG message will show on the display as calibration is
failed.


6.6 How to do the Gain Calibration via CH2

The Offset Calibration via CH2 is for checking DC +5V. Please do the
following procedures to finish the calibration.

1. Press UTILITY > CAL > Enter Security Code > Cal Diag > Gain

Inp2.

2. Use the BNC T-connector with 2 BNC cords to link an Agilent

33220A and a DMM (multimeter).

3. Under the Agilent 33220A operation, the procedures should be

Utility > DC ON > Output Setup > Hi-Z.

4. Adjust the voltage on the Agilent 33220A, and let the DMM

displays +5.000V ±0.001V.

5. Remove the input source, from the DMM, on the counter.

6.

Press the softkey under RunCal to do the calibration. Then, the
counter’s display will show “Calibrating…”. The finished message

Advertising