softkey under Save.). For more calibration information, please check
the List 4-1.

5. Press the softkey under Run Cal when everything is set ready.

Figure 4-27 Figure 4-28

Item

Softkeys

Description

1. Gain Inp1 + Run Cal

DC +5V Calibration at CH1

2. Gain Inp2 + Run Cal

DC +5V Calibration at CH2

3. Gain Att1 + Run Cal

DC +9V ATT x 10 Calibration at CH1

4. Gain Att2 + Run Cal

DC +9V ATT x 10 Calibration at CH2

5. Offs Inp1 + Run Cal

DC +0V Calibration at CH1

6. Offs Inp2 + Run Cal

DC +0V Calibration at CH2

7. Offs Att1 + Run Cal

DC +0V ATT x 10 Calibration at CH1

8. Offs Att2 + Run Cal

DC +0V ATT x 10 Calibration at CH2

9. T-F Quick + Run Cal

Time Domain Calibration (incl.

External Arm)

10. T-F Fine1 + Run Cal

Time Interval Calibration

11. T-F Fine2 + Run Cal

Time Interval Calibration ( incl.

External Arm)

12. T-F Freq + Run Cal

10 MHz Calibration

13.

Save

Save the settings in the memory

List 4-1

Remote Interface Command
Use these commands to run calibration:

:CALibration:SECurity:STATe

<Boolean>,<NRf>

System Setup
Under the System Setup menu as shown in the Figure 4-29, users can
run self-test, adjust

the beeper setting, choose dot convention (i.e.,

using comma or period to separate the integral portion and fraction
portion of a number), and TOUT (i.e., If the time out “ON” is chosen, the
time out mode will be enabled, and as measurement time exceeds