BNC 1105 Universal Counter User Manual

Page 137

Advertising
background image

137

137

137

137

Queries the current measurement results of the :SENSe commands (no scale or offset

applied).

Return

Result is formatted according to :FORMat[:DATA] ASCii|REAL command.

Notes

If using the ASCii format, numeric data is transferred as ASCII bytes in

<NR3> format. The number of significant digits ranges from 1 to 10

depending on the resolution.

If there is no valid measurement result, Not A Number 9.91E37 will be

returned and error -230 will be generated.

If this command is executed while a measurement is in progress, the prior

measurement result will be returned.

If this command is executed while an automatically armed Totalize

measurement is in progress, Not A Number 9.91E37 will be returned and

error -221 will be generated. User should send :ABORt command to abort

the mea

surement so that the error can be avoided.

[:SENSe]EVENt2:FEED “[:]INPut[1]|[:]INPut2”

Sets or queries the state of the common/separate.

Return

The string “INP” or “INP2” is returned.

Notes

This will be set to its default, which is “INP2”, after the counter has been turned off or

after a remote interface reset.

All functions but Time Interval allow fixed input route setting.

[:SENSe]EVENt[1|2]:HYSTeresis:RELative <numeric value> [PCT]

Sets or queries the size of the hysteresis window as a percentage of the allowable

hysteresis. For example, 0% is for the minimum hysteresis setting and 100% is for the

maximum. Specifying 100% or MAXimum provides the greatest noise immunity (lowest

sensititity), while specifying 0% or MINimum provides the least noise immunity.