The Channel 1 and Channel 2 share the same trigger setting so that the measurement channel

has auto-trigger enabled at 50% with a positive slope.

<source list>

Default

(@1),(@2)

Range

(@1),(@2)

:MEASure[:SCALar][:VOLTage]:PTPeak? [(@1)|(@2)]

Makes Peak-to-Peak measurement.

<source list>

Default

(@1)

Range

(@1)|(@2)

:MEASure[:SCALar][:VOLTage]:CENTer? [(@1)|(@2)]

Measures center voltage.

<source list>

Default

(@1)

Range

(@1)|(@2)

:MEASure[:SCALar][:VOLTage]:TINTerval? [(@1),(@2)]

Measures Time Interval.

The first channel is the start channel and the second channel is the stop channel.

The Channel 1 and Channel 2 share the same trigger setting so that the measurement channel

has auto-trigger enabled at 50% with a positive slope.

The measurement arming is set to “auto” with no delay. The input routing is coupled to separate

inputs.

<source list>

Default

(@1),(@2)

Range

(@1),(@2)

:MEASure[:SCALar][:VOLTage]:TOTalize:TIMed? [<gate_time>][,(@1)]

Makes Totalize measurement during the specified <gate_time>.

The trigger settings are coupled so that the measurement channel has auto-trigger

disabled with a positive slope.

<gate time>

Default

0.1 sec

Range

1E-3 to 1,000 sec

Resolution

1E-5 sec

<source list>

Default

(@1)

Range

(@1)

:MEMory:NSTates?

Queries the number of available *SAV/*RCL states in the counter.