Sets or queries the percentage of the peak-to-peak range of the signal at which the

instrument automatically triggers. The range is from 0 to 100 PCT, and the resolution is

10 PCT.

The counter automatically measures and computes the trigger level according to the

percentage specified by the command :[:SENS]:EVEN[1︱2]:LEV:REL when it receives

this command.

Return

Numeric data transferred as ASCII bytes in <NR1> format.

Notes

This will be reset to 50 PCT after the counter has been turned off or a remote

interface reset.

This command only applies when [:SENS]:EVEN[1

︱2]:LEV[:ABS]:AUTO is ON

[:SENSe]EVENt[1|2]:SLOPe POSitive|NEGative

Sets or queries which edge of the input signal will be considered an event when the

counter measures frequency, period, frequency ratio, time interval, totalize or phase.

If specifying the positive slope, a signal ascending from a lower voltage to a higher

voltage will define an event at the upper hysteresis limit, on the other hand, if specifying

the negative slope, a signal descending from a higher voltage to a lower voltage will

define an event at the lower hysteresis limit.

Return

A sequence of ASCII-encoded bytes, POS or NEG.

Notes

This will be set to Positive after the counter has been turned off or a remote interface

reset.

The slope specified here is irrelevant from to the ones used in Rise/Fall Time, Duty

Cycle and Pulse Width measurements.

[:SENSe]EVENt3:LEVel[:ABSolute]?

Queries the channel 3 trigger level.

Return

Numeric data transferred as ASCII bytes in <NR3> format with six significant digits