digits range from 1 to 15, depending on the resolution.

Notes

The statistics operation should be enabled (:CALC3:AVER[:STATe] ON ) before using

this command. If not, Not A Number 9.91E37 will be returned and error -221 will be

generated.

The result is separated by commas, and the display order is mean value, standard

deviation, minimum value and then maximum value.

If there is no valid result, Not A Number 9.91E37 will be returned and error -230 will

be generated.

If the current measurement operation is Totalize or Voltage Peaks, Not A Number

9.91E37 will be returned and error -221 will be generated.

:CALCulate3:AVERage:CLEar

Clears the statistics data, sets the statistics counts to zero and reports the negative

status condition to bit 8 of the Operation Status Register.

Notes

If the statistics operation is not enabled (:CALC3:AVER[:STATe] OFF, error -221 will

be generated.

:CALCulate3:AVERage:COUNt <numeric value>

Sets or queries the number of measurement that to be used in the statistics operation.

The range of this number is from 2 to 1,000,000.

Return

Numeric data transferred as ASCII bytes in <NR1> format.

Notes

This will be set to the counter’s default, 100, after the counter has been turned off or

a remote interface reset.

When the specified number of measurement has been collected, another set of

measurements has to be collected before the counter does another statistics

calculation.

:CALCulate3:AVERage:COUNt:CURRent?

Queries the number of measurement collected so far for statistics calculation. The range

is from 0 to 1,000,000.