Error messages, 1 command errors – BNC 1105 Universal Counter User Manual

Page 88

Advertising
background image

88

88

88

88

7

Error Messages


Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. Use :SYST:ERR? to read errors.
When user has read all errors from the queue, the error queue becomes
empty and any error query will return +0, “No error”. The MODEL 1105
beeps once each time an error occurs.

Should more than 29 errors have existed, the last error stored in the
queue (the most recent error) is replaced with -350, “Queue overflow”.
No additional errors are saved until you remove errors from the queue.

The error queue is cleared when power is off or after a *CLS (clear status)
command is executed. The *RST (reset) command does not clear the
error queue.

7.1 Command Errors

+0 No error

There is no error in the error queue. (:SYSTem:ERRor? query) or the queue

was cleared by power-on or *CLS.

-100 Command error

This is the generic syntax error used when the Counter cannot detect more

specific errors.

-101 Invalid character

A syntactic element contains a character that is invalid for that type. For

example, a header containing an ampersand, :INP:COUP& AC.

-102 Syntax error

An unrecognized command or data type was encountered.

-103 Invalid separator

The parser was expecting a separator but encountered an illegal character.

-104 Data type error

The parser recognized a data element different than one with the allowed

Advertising