BNC 1105 Universal Counter User Manual

Page 145

Advertising
background image

145

145

145

145

Sets or queries the start arm for Phase measurements.

Return

A sequence of ASCII-encoded bytes. IMM means Immediate and EXT means

External.

Note

This will be set to Immediate after the counter has been turned off or a remote

interface reset.

[:SENSe]:ROSCillator:EXTernal:FREQuency?

Under the timeBase measurement, (Fno) will be shown on the counter display if there is

no inputted frequency, or (Fxx) will be shown on the counter and “Not a number” on the

PC side if the inputted frequency is an invalid value. On the other hand, the command

ROSCillator:SOURce

will reply “Not a number” under the INTernal condition, and

“1.00000E+007” under the EXTernal condition.

Return

Numeric data transferred as ASCII bytes in <NR3> format with six significant digits.

Note

If the current reference timebase is internal or if the current reference timebase is

external, but the frequency is not known, Not A Number 9.91E37 will be returned.

[:SENSe]:ROSCillator:SOURce INTernal|EXTernal

Sets or queries the current reference timebase.

Return

A sequence of ASCII-encoded bytes. INT means that the timebase is internal

reference. EXT means that external.

Notes

Specifying the reference timebase automatically disables

[:SENS]:ROSC:SOUR:AUTO ([:SENS]:ROSC:SOUR:AUTO OFF).

When [:SENS]:ROSC:SOUR:AUTO ON, this query tells which timebase reference is

used.

Save/recall operation does not apply on this value.