BNC 1105 Universal Counter User Manual

Page 134

Advertising
background image

134

134

134

134

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

<source list>

default

(@1)

range

(@1)

:MEASure[:SCALar][:VOLTage]:PERiod? [<expected_value> [,<resolution>]]

[,(@1)|(@2)|(@3)]

Sets the measurement arming to “digits” and measures the period.

The <expected value> and <resolution> parameters are used to configure the number of digits

of resolution arming setting.

The Channel 1 and Channel 2 share the same trigger setting so that the measurement channel

has auto-trigger enabled at 50% with a positive slope.

<expected value>

Ch1, Ch2

Default

100 ns

Range

2 ns ~ 10 sec

Resolution

<expected value> should be within 10% of the input

period for optimum arming configuration

<expected value>

Ch3

Default

2 ns

Range

0.17 ns ~ 2.67 ns

Resolution

<expected value> should be within 10% of the input

period for optimum arming configuration

<resolution>

Ch1, Ch2

Default

Value which indicates 4 digits of resolution for the

specified <expected value>

Range

1E-23 ~ 1E-2 sec (15 to 3 digits of resolution)

Resolution

<resolution> should use a mantissa of 1.0 and be an

even power of 10

<resolution>

Ch3

Default

Value which indicates 4 digits of resolution for the

specified <expected value>

Range

1E-24 ~ 1E-11 sec (15 to 3 digits of resolution)

Resolution

<resolution> should use a mantissa of 1.0 and be an

even power of 10

<source list>

Default

(@1)

Range

(@1)(@2)(@3)

:MEASure[:SCALar][:VOLTage]:PHASe? [(@1),(@2)]

Sets the measurement arming to “auto” and measures Phase.

Advertising