A single ASCII-encoded byte. 0 means AUTO OFF and 1 means AUTO ON.

Note

This will be set back to AUTO OFF after *RST command.

:INITiate:CONTinuous <Boolean>

Sets or queries the enable for initiating measurements continuously.

When the CONTinuous is set to ON, the measurements are initiated continuously,

when the CONTinuous is set to OFF, the counter does not issue any

measurement until it receives :INITiate[:IMMediate] command or CONTinuous

is set to ON.

Return

A single ASCII-encoded byte. 0 means CONTinuous OFF and 1 means

CONTinuous ON.

Notes

This will be set back to CONTinuous OFF after *RST command.

When :INIT:CONT ON command is sent, the counter

o

Invalidates the statistics results,

o

Clears the current statistics count to 0,

o

Reports the negative status condition (NOT computing statistics) to bit 8

of Operation Status Register.

When :CALC2:LIM:CLE:AUTO is ON and whenever the counter

receives :INIT:CONT ON command, it :

o

Invalidates the limit data,

o

Clears the pass and fail counts (:CLAC2:LIM:FCO and :CALC2:LIM:PCO)

in the limit testing.

o

Sets the Limit-Detect output to the in-limit voltage level.

When [:SENSe]:EVEN:LEV[:ABS]:AUTO is ON and :INIT:CONT ON

command is sent, the counter issues an auto-trigger on the source channel

and at the beginning of each measurement cycle while :INIT CONT is ON.

The state of :TRIG:COUN:AUTO has no affect on the operation

of :INIT:CONT ON.

:INIT:CONT ON operates as if :TRIG:COUN was 1.