10 utility operations – BNC 1105 Universal Counter User Manual

Page 58

Advertising
background image

58

58

58

58

To Power on Recall
1. Press Save & Recall button.
2. Press the softkey under Save to select the memory number from 1

to 20 for stocking present settings. As well the arrow and numeric
keys are available to be used. (It’s better that users take memo on
each saved number.)

3. Press the softkey under Done to finish.
4. Press the softkey under RwrOn Recall to toggle between the saved

memory numbers shown on the display.

5. Press the softkey under Done to select the desire number as the

power-on default.

4.10 Utility Operations

The Utility operations let users execute manual calibration, configure
interface, enable or disable Beep function, run self-test, and set up
timebase as shown in the Figure 4-26.

Figure 4-26

I/O configuration
Refer to chapter 5 for interface configuration.

Calibration
Users need a security code. Only when a correct security code is entered,
users will have access to the calibration menu.

1. Press the UTILITY button, and it works with a lighting signal.
2. Use the softkey under CAL to select the calibration operation.
3. Use the numeric keypad or the arrow keys to enter the correct

security code as shown in the Figure 4-27.

4. Press the softkey under Cal Diag to manually set calibration

conditions, such as Gain, Offs and T-F as shown in the Figure 4-28
(users can save the settings according to purposes by pressing the

Advertising