A sequence of ASCII-encoded bytes(“SENS”*SCALE+OFFSET) terminated by a new

line and EOI.

Note

It should be the last query in a terminated program message or error-440 will be

generated.

:CALCulate[1]:MATH[:EXPRession]:NAME SCALE_OFFSET

Or

:CALCulate[1]:MATH[:EXPRession]:SELect SCALE_OFFSET

Defines the name of the expression used in mathematic operation.

Return

A sequence of ASCII-encoded bytes:SCALE_OFFSET.

:CALCulate[1]:MATH:STATe <Boolean>

Enables or disables the mathematic operation.

Return

An ASCII-encoded byte. 0 means OFF, and 1 means ON.

Note

This command will clear the limit counts(:CALC:LIM:FCO, :CALC:LIM:PCO).

The command :CALCulate2

This group performs limit testing and data transfer.

:CALCulate2:FEED “[:]CALCulate[1]”

Sets or queries the data flow to be fed into the CALCulate2 block.

Return

The string “CALC” is returned.

:CALCulate2:IMMediate

Instructs the counter to re-compute the current data without re-acquiring data.

This command will not affect: