5 counter-specific errors – BNC 1105 Universal Counter User Manual

Page 93

Advertising
background image

93

93

93

93

7.5 Counter-Specific Errors

+2000 Offset calibration on channel 1 failed

+2001 Offset calibration on channel 2 failed

+2002 Gain calibration on channel 1 failed

+2003 Gain calibration on channel 2 failed

+2004 Interpolator calibration failed

+2005 Oscillation calibration failed

+2006 Time interval calibration failed

+2007 Measurement hardware calibration failed

A measurement calibration failed on the last measurement.

+2009 No valid external timebase

Users have selected external reference and there is no external reference applied to

the rear-panel Ref In connector, or the external signal is not an allowed frequency.

+2010 External timebase failed during measurement

Users have selected the auto reference mode and the Counter detected that the

external reference became invalid during the measurement. Therefore, the current

result is not valid, and the Counter switches to using the internal reference.

+2011 Recall setup failed; hardware failure

*RCL failed.

+2011 Recall setup failed; empty register

*RCL failed because the register specified for recall is empty.

+2012 Save setup failed

*SAV failed, or *RCL couldn’t save to register 0.

+2013 EEPROM failed

A request to update a setting which is stored in the EEPROM

:SYST:COMM:SER:CONT:DTR, :DISP[:WIND]:TEXT:RAD,

:CAL[:DATA] resulted in a hardware failure.

Advertising