BNC 1105 Universal Counter User Manual

Page 97

Advertising
background image

97

97

97

97

Channel 1 and 2

Trigger

Default setting is Auto Trigger at 50 %

“Auto” Gate Time

0.1 sec

STD CH 3

375 MHz to 6 GHz (0.166 ns to 2.6 ns)

Frequency Ratio

CH 1/ CH 2, CH 1/ CH 3, CH 2/ CH 1, CH 3/ CH1

(Measurement is specified over the full signal range of each input.)

Results Range

10E-10 to 10E+11

“Auto” Gate Time

0.1 sec

Time Interval

Measurement is specified over the full signal ranges of Channels 1 and 2. The width of the

pulse must be greater than 1 ns, frequency range to 300 MHz.

Trigger

Default setting is Auto Trigger at 50 %

Results Range

-0.5 ns to 10E+5 s

Resolution

40 ps

RMS Resolution

120 ps

Systematic

Uncertainty

±(TI × Time Base Error) ± Trigger Level Timing Error ± 500 ps Differential Channel Error

Pulse Width Time

Measurement is specified over the full signal range of Channel 1. The width of the pulse

must be greater than 1 ns frequency range to 300 MHz).

Pulse Selection

Positive or Negative

Trigger

Default setting is Auto Trigger at 50%

Results Range

1.5 ns to 10E+5 s

Resolution

40 ps

RMS Resolution

120 ps

Systematic

Uncertainty

± (Pulse Width Time x Time Base Error) ± Trigger Level Timing Error ± 500 ps Differential

Channel Error.

Rise/Fall Time

Measurement is specified over the full signal range of Channel 1. The width of the pulse

must be greater than 1 ns frequency range to 300 MHz).

Edge Selection

Positive or Negative

Trigger

Default setting is Auto Trigger at 10% and 90%

Results Range

2 ns to 10E+5 s

Resolution

40 ps

RMS Resolution

120 ps

Systematic

Uncertainty

± (Edge Time x Time Base Error) ± Trigger Level Timing Error ± 500 ps Differential

Channel Error

Phase

Measurement is specified over the full signal range of each input. The width of the pulse

must be greater than 1 ns, frequency range to 300 MHz

Results Range

-180° to +360°

Resolution

40 ps

RMS Resolution

120 ps

Systematic

Uncertainty

± (Trigger Level Timing Error) ×Frequency

Duty Cycle

Measurement is specified over the full signal range of Channel 1. The width of the pulse

must be greater than 1 ns, frequency range to 300 MHz

Pulse Selection

Positive or Negative

Trigger

Default setting is Auto Trigger at 50 %

Results Range

0 to 1

Resolution

40 ps

RMS Resolution

120 ps

Systematic

Uncertainty

± Trigger Level Timing Error ± 500 ps Differential Channel Error

Totalize

Measurement is specified over the full signal range of Channel 1. The width of the pulse

must be greater than 1 ns, frequency range to 400 MHz

Pulse Selection

Positive or Negative

Advertising