BNC 1105 Universal Counter User Manual

Page 130

Advertising
background image

130

130

130

130

:MEASure[:SCALar][:VOLTage]:FALL:TIME?

[<lower_reference> [,upper_reference>]][,(@1)]

Sets the measurement arming to “auto” and measures Fall Time with the specified

parameters.

The <lower_reference> and <upper_reference> are used to specify the lower and upper

points on the falling edge of the input signal. They can be specified as either a percentage

or an absolute voltage.

The <lower_reference> and <upper_reference> are used to configure the trigger setting.

When they are in percentage or defaults, auto-trigger is enabled and trigger level is the

specified value. When they are specified in voltage, auto-trigger is disabled and the

trigger level is the specified value.

<lower_reference>

default

10 pct

Percent range

0 to 100 pct

Percent resolution

10%

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

<upper_reference>

default

90 pct

Percent range

0 to 100 pct

Percent resolution

10%

Voltage range

-5.125V to 5.125V for ×1 attenuation

-51.25V to 51,25V for ×10 attenuation

Voltage resolution

0.0025V for ×1 attenuation
0.025V for ×10 attenuation

<source list>

default

(@1)

range

(@1)

:MEASure[:SCALar][:VOLTage]:RISE:TIME?

[<lower_reference> [,upper_reference>]][,(@1)]

Sets the measurement arming to “auto” and measures Rise Time with the specified

parameters.

The <lower_reference> and <upper_reference> are used to specify the lower and upper

points on the falling edge of the input signal. They can be specified as either a percentage

or an absolute voltage.

The <lower_reference> and <upper_reference> are used to configure the trigger setting.

When they are in percentage or defaults, auto-trigger is enabled and trigger level is the