152

152

152

152

reset.

When the staring and the stopping points are both set to immediate, the

command :INIT[:IMM] can be used to start totalizing and the command :ABORt can

be used to end it. Before the “auto” totalize measurement is terminated, no totalize

result will be available.

[:SENSe]:TOTalize:ARM:STOP:TIMer <numeric value> [S]

Sets or queries the gate time in arming Totalize measurements. The range is 1E-3 to

1,000 seconds and the resolution is 0.01E-3 seconds.

Return

Numeric data transferred as ASCII bytes in <NR3> format with six significant digits.

Notes

This will be set to 100E-3 seconds after the counter has been turned off or a remote

interface reset.

This command only applies when [:SENSe]:TOT:ARM:STOP:SOUR TIM is used.

[:SENSe]:VOLTage:MODE FAST

|SLOW

To set or query “FAST” or “SLOW” under the Mode of Voltage Peak, the “FAST”

is recommended if the input source is over 10 KHz, and the “SLOW” is

recommended if the input source is over 100 Hz.

Return

A sequence of ASCII-encoded bytes, FAST or SLOW.

The Commands :SYSTem

These commands deal with the capabilities that have no direct effects on
the counter’s performance.

:SYSTem:COMMunicate:SERial:SOURce USB

|DB9

Sets or queries the hardware pacing scheme. As the LIMIT PASS/FALL is set to USB, the

output port will be via USB. As the LIMIT PASS/FALL is set to DB9, the output port will be

via 9-pin plug like RS-232.

Return