AMETEK ReFlex User Manual

Page 136

Advertising
background image

ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-51

14. Program to DAC to 0.5% above the middle range

CAL<n>:OUTP:DAC:COUNTS 33080

15. Wait for the DC voltmeter to stabilize. Make an accurate measurement and then

enter the DC voltage using following command

CAL<n>:OUTP:DAC:POINT2 <value-2>

16. Perform the DAC Calibration

CAL<n>:OUTP:DAC:OFFSET:CALC

17. Measure the DC output voltage, if the voltage is not 0.0 ± 10mVolts, read the

DAC offset value with CAL<n>:OUTP:DAC:OFFSET?, and then using the value
returned (a Hex number) as a starting point, “walk” the DC voltage toward 0.0
volts by increasing the value if the DC value is negative, and decreasing the
value if it is positive. One count is

≈ 10mVolts. The command to change the

DAC offset value is CAL<n>:OUTP:DAC:OFFSET <value>.

18. Turn the DC correction loop back on

CAL<n>:OUTP:DCLOOP ON

19. Turn off the unit

OUTP<n>:STATE 0

20. Send the system error query command to make sure calibration was successful.

SYST:ERR?

21. If there is no error returned at the system error query command then save the

calibration result with the following command

CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK

22. Reset the unit

*RST<n>