17 ldf calibration for voltage range high – AMETEK ReFlex User Manual
Page 151

AC Power Supplies
ReFlex Power™
4-66
M380056-01 Rev K
4.6.17
LDF
C
ALIBRATION FOR
V
OLTAGE
R
ANGE
H
IGH
The following procedure calibrates the Line drop fault supervisory.
1. Ensure that no load is connected to the output.
2. Issue clear and reset commands:
*CLS
*RST<n>
3. Set the voltage range of the power supply to high
SOUR<n>:VOLT:RANGE 1
4. Turn the unit on
OUTP<n>:STATE 1
5. Program the output voltage to count 17000
CAL<n>:OUTP:VOLT:COUNTS 17000
6. Send the calibration command to start the internal calibration
CAL<n>:OUTP:LDF:CALC
7. Turn the unit off
OUTP<n>:STATE 0
8. Send the system error query command to make sure calibration was successful.
SYST:ERR?
9. If there is no error returned at the system error query command then save the
calibration result with the following commands
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
10. Optional Commands:
Query the voltage calibration points:
CAL<n>:OUTP:LDF:GAIN ?
11. Reset the unit
*RST<n>