7-61
7. Send the calibration command to calculate the gain and offset of the
low range current calibration
CAL<n>:INP:CURR:CALC
8. Send the system error query command to make sure calibration was
successful.
SYST:ERR?
9. If there is no error returned at the system error query command then
save the calibration result with the following command
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
10. Reset the unit
*RST<n>