Active Loads

ReFlex Power™

7-58

M380056-01 Rev J

7. Set the DAC count to 50.0% of its full range

CAL<n>:INP:CURR:COUNTS 32000

8. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL<n>:INP:CURR:FIVEPOINT3 <value-3>

9. Set the DAC count to 72.5% of its full range

CAL<n>:INP:CURR:COUNTS 43000

10. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL<n>:INP:CURR:FIVEPOINT4 <value-4>

11. Set the DAC count to 95% of its full range

CAL<n>:INP:CURR:COUNTS 56000

12. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL<n>:INP:CURR:FIVEPOINT5 <fval>

13. Set the DAC to zero counts

CAL<n>:INP:CURR:COUNTS 0