AMETEK ReFlex User Manual

Page 348

Advertising
background image

ReFlex Power™

Active Loads

M380056-01 Rev J

7-73

7. After DC voltage stabilizes. Enter the voltage using SCPI

commands

CAL<n>:INP:VOLT 2 <value-2>

8. After DC current stabilizes. Enter the current using SCPI commands

CAL<n>:INP:CURR 2 <value-2>

9. Send the calibration command to calculate the gain and offset

CAL<n>:INP:COND:CALC

10. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

11. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

12. Reset the unit

*RST<n>