LPAL

V

OLTAGE

C

ALIBRATION

PART

III

OF

V

This procedure calibrates the voltage for voltage gain 2 and current range
high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.

This procedure requires maximum voltage 218V and maximum current 1.5
Amps.

1. Turn the active load on

INP<n>:STATE 1

2. Set the conductance count to 0x1999

CAL<n>:INP:COND:COUNTS 0x1999

3. Set the input DC voltage to 54 Volts. After the input DC voltage

stabilizes enter the first exact voltage using following command

CAL<n>:INP:VOLT 1 <value-1>

4. Set the input DC voltage to 218 Volts. After the input DC voltage

stabilizes enter the second exact voltage using following command

CAL<n>:INP:VOLT 2 <value-2>

5. Send the calibration command to calculate the gain and offset:

CAL<n>:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful.

SYST:ERR?