AMETEK ReFlex User Manual

Page 141

Advertising
background image

AC Power Supplies

ReFlex Power™

4-56

M380056-01 Rev K

4.6.10

C

URRENT

ADC

O

FFSET

C

ALIBRATION FOR

V

OLTAGE

R

ANGE

L

OW

The following procedure calibrates offset of the current ADC for the voltage range
low.

1. Ensure that no load is connected to the output.

2. Issue clear and reset commands:

*CLS
*RST<n>

3. Set the voltage range of the power supply to low

SOUR<n>:VOLT:RANGE 0

4. Perform the voltage ADC offset calibration

CAL<n>:MEAS:IADC:OFFSET:CALC

5. Send the system error query command to make sure calibration was successful.

SYST:ERR?

6. If there is no error returned at the system error query command then save the

calibration result with the following commands

CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK

7. Optional Commands:

To query the current ADC offset:

CAL<n>: MEAS:IADC:OFFSET ?

8. Reset the unit

*RST<n>