AMETEK ReFlex User Manual
Page 141

AC Power Supplies
ReFlex Power™
4-56
M380056-01 Rev K
4.6.10
C
URRENT
ADC
O
FFSET
C
ALIBRATION FOR
V
OLTAGE
R
ANGE
L
OW
The following procedure calibrates offset of the current ADC for the voltage range
low.
1. Ensure that no load is connected to the output.
2. Issue clear and reset commands:
*CLS
*RST<n>
3. Set the voltage range of the power supply to low
SOUR<n>:VOLT:RANGE 0
4. Perform the voltage ADC offset calibration
CAL<n>:MEAS:IADC:OFFSET:CALC
5. Send the system error query command to make sure calibration was successful.
SYST:ERR?
6. If there is no error returned at the system error query command then save the
calibration result with the following commands
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
7. Optional Commands:
To query the current ADC offset:
CAL<n>: MEAS:IADC:OFFSET ?
8. Reset the unit
*RST<n>