13. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL<n>:OUTP:VOLT:FIVEPOINT2 <value-2>
14. Program the output voltage to 50% of the full scale voltage
CAL<n>:OUTP:VOLT:COUNTS 9649
15. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL<n>:OUTP:VOLT:FIVEPOINT3 <value-3>
16. Program the output voltage to 72.5% of the full scale voltage
CAL<n>:OUTP:VOLT:COUNTS 13990
17. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL<n>:OUTP:VOLT:FIVEPOINT4 <value-4>
18. Program the output voltage to 95% of the full scale voltage
CAL<n>:OUTP:VOLT:COUNTS 18332
19. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL<n>:OUTP:VOLT:FIVEPOINT5 <value-5>
20. Perform the voltage calibration using following command
CAL<n>:OUTP:VOLT:CALC
21. Send the system error query command to make sure calibration was successful.
SYST:ERR?
22. If there is no error returned at the system error query command then save the
calibration result with the following command
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
23. Optional Commands:
Query the voltage calibration points:
CAL<n>:OUTP:VOLT:FIVEPOINT ?
24. Reset the unit