ReFlex Power™

Active Loads

M380056-01 Rev J

7-65

7. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

8. Reset the unit

*RST<n>