AMETEK ReFlex User Manual
Page 350
Advertising

ReFlex Power™
Active Loads
M380056-01 Rev J
7-75
7. After DC voltage stabilizes. Enter the voltage using SCPI
commands
CAL<n>:INP:VOLT 2 <value-2>
8. After DC current stabilizes. Enter the current using SCPI commands
CAL<n>:INP:CURR 2 <value-2>
9. Send the calibration command to calculate the gain and offset
CAL<n>:INP:COND:CALC
10. Send the system error query command to make sure calibration was
successful.
SYST:ERR?
11. If there is no error returned at the system error query command then
save the calibration result with the following command
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
12. Reset the unit
*RST<n>