Active Loads

ReFlex Power™

7-48

M380056-01 Rev J

CAL<n>:INP:VOLT 2 <value-2>

8. After DC current stabilizes, enter the current:

CAL<n>:INP:CURR 2 <value-2>

9. Send the calibration command to calculate the gain and offset:

CAL<n>:INP:COND:CALC

10. Send the system error query command to make sure calibration was

successful:

SYST:ERR?

11. If there is no error returned at the system error query command,

then save the calibration result as follows:

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

12. Reset the unit:

*RST<n>