25 lpal voltage calibration – part iv of v, Lpal – AMETEK ReFlex User Manual
Page 343

Active Loads
ReFlex Power™
7-68
M380056-01 Rev J
7.6.25
LPAL
V
OLTAGE
C
ALIBRATION
–
P
ART
IV
OF
V
This procedure calibrates the voltage for voltage gain 1 and current range
high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.
This procedure requires maximum voltage 500.0 V and maximum current
0.5 Amps.
1. Turn the active load on
INP<n>:STATE 1
2. Set the conductance count to 0x418
CAL<n>:INP:COND:COUNTS 0x418
3. Set the input DC voltage to 10 Volts. After the input DC voltage
stabilizes enter the first exact voltage using following command
CAL<n>:INP:VOLT 1 <value-1>
4. Set the input DC voltage to 500 Volts. After the input DC voltage
stabilizes enter the second exact voltage using following command
CAL<n>:INP:VOLT 2 <value-2>
5. Send the calibration command to calculate the gain and offset:
CAL<n>:MEAS:OFFSET:CALC
6. Send the system error query command to make sure calibration was
successful.