AMETEK ReFlex User Manual
Page 159

AC Power Supplies
ReFlex Power™
4-74
M380056-01 Rev K
15. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR2 <value-2>
16. Set the output frequency to 1000Hz
SOUR<n>:FREQ 1000
17. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR3 <value-3>
18. Set the output frequency to 2000Hz:
SOUR<n>:FREQ 2000
19. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR4 <value-4>
20. Set the output frequency to 3000Hz:
SOUR<n>:FREQ 3000
21. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR5 <value-5>
22. Set the output frequency to 4000Hz:
SOUR<n>:FREQ 4000
23. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR6 <value-6>
24. Set the output frequency to 5000Hz:
SOUR<n>:FREQ 5000
25. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:FREQ:CURR7 <value-7>
26. Perform the voltage calibration:
CAL<n>:OUTP:FREQ:CALC
27. Send the system error query command to make sure calibration was successful:
SYST:ERR?