AMETEK ReFlex User Manual

Page 149

Advertising
background image

AC Power Supplies

ReFlex Power™

4-64

M380056-01 Rev K

CAL<n>:OUTP:VOLT:FIVEPOINT2 <value-2>

12. Program the output voltage to 50% of the full scale voltage

CAL<n>:OUTP:VOLT:COUNTS 9649

13. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL<n>:OUTP:VOLT:FIVEPOINT3 <value-3>

14. Program the output voltage to 72.5% of the full scale voltage

CAL<n>:OUTP:VOLT:COUNTS 13990

15. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL<n>:OUTP:VOLT:FIVEPOINT4 <value-4>

16. Program the output voltage to 95% of the full scale voltage

CAL<n>:OUTP:VOLT:COUNTS 18332

17. Wait for AC voltage to stabilize. Measure the output voltage using the AC

voltmeter. Enter the measured voltage using following command

CAL<n>:OUTP:VOLT:FIVEPOINT5 <value-5>

18. Perform the voltage calibration using following command

CAL<n>:OUTP:VOLT:CALC

19. Send the system error query command to make sure calibration was successful.

SYST:ERR?

20. If there is no error returned at the system error query command then save the

calibration result with the following command

CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK

21. Optional Commands:

Query the voltage calibration points: CAL<n>:OUTP:VOLT:FIVEPOINT ?

22. Reset the unit

*RST<n>