AMETEK ReFlex User Manual

Page 310

Advertising
background image

ReFlex Power™

Active Loads

M380056-01 Rev J

7-35

12. Wait for the DC current to stabilize. After the current reading is stabilized,

enter the current using following command:

CAL<n>:INP:CURR:FIVEPOINT5 <fval>

13. Set DAC counts to zero:

CAL12:INP:CURR:COUNTS 0

14. Send calibration command to make sure that all the DAC counts and

current values are valid:

CAL<n>:INP:CURR:CALC

15. Send the system error query command to make sure calibration was

successful:

SYST:ERR?

16. If there is no error returned at the system error query command then save

the calibration result with the following commands:

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

17. Reset the unit:

*RST<n>

Advertising