14. Send calibration command to make sure that all the DAC counts

and current values are valid

CAL<n>:INP:CURR:CALC

15. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

16. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

17. Reset the unit

*RST<n>