14 hpal resistance calibration – part ii of v, Hpal – AMETEK ReFlex User Manual

Page 320

Advertising
background image

ReFlex Power™

Active Loads

M380056-01 Rev J

7-45

7.6.14

HPAL

R

ESISTANCE

C

ALIBRATION

P

ART

II

OF

V

This procedure calibrates the resistance for voltage gain 4 and current
range high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Connect the
sense positive-to-positive lead and sense negative-to-negative lead. Also
connect the shunt and the DC current meter between the active load and
DC power supply

This procedure requires maximum voltage 20.0 V and maximum current
3.0Amps.

1. Set the DC input voltage to 20 Volts

2. Turn the active load on:

INP<n>:STATE 1

3. Set the conductance count to 0x1FFFF.

CAL<n>:INP:COND:COUNTS 0x1FFFF

4. After DC voltage stabilizes. Enter the voltage using SCPI

commands

CAL<n>:INP:VOLT 1 <value-1>

5. After DC current stabilizes. Enter the current using SCPI commands

CAL<n>:INP:CURR 1 <value-1>

6. Set the conductance count to 0xF0F0.

CAL<n>:INP:COND:COUNTS 0x3CF3C