7-41

7.6.11

HPAL

V

OLTAGE

C

ALIBRATION

P

ART

IV

OF

V

This procedure calibrates the voltage for voltage gain 1 and current range high.
Make sure that shunt and the DC voltmeter is connected between the positive
and negative lead of the active load. Also connect the sense positive-to-positive
lead and sense negative-to-negative lead.

This procedure requires maximum voltage 500.0V and maximum current 1.0
Amps

1. Turn the active load on:

INP<n>:STATE 1

2. Set the conductance count to 0x831:

CAL<n>:INP:COND:COUNTS 0x831

3. Set the input DC voltage to 10 Volts. After the input DC voltage

stabilizes enter the first exact voltage:

CAL<n>:INP:VOLT 1 <value-1>

4. Set the input DC voltage to 500 Volts. After the input DC voltage

stabilizes enter the second exact voltage:

CAL<n>:INP:VOLT 2 <value-2>

5. Send the calibration command to calculate the gain and offset:

CAL<n>:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful:

SYST:ERR?

7. If there is no error returned at the system error query command then

save the calibration result as follows:

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

8. Reset the unit

*RST<n>