AMETEK ReFlex User Manual
Page 325
Advertising

Active Loads
ReFlex Power™
7-50
M380056-01 Rev J
7. After DC voltage stabilizes, enter the voltage:
CAL<n>:INP:VOLT 2 <value-2>
8. After DC current stabilizes, enter the current:
CAL<n>:INP:CURR 2 <value-2>
9. Send the calibration command to calculate the gain and offset:
CAL<n>:INP:COND:CALC
10. Send the system error query command to make sure calibration was
successful:
SYST:ERR?
11. If there is no error returned at the system error query command,
then save the calibration result as follows:
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
12. Reset the unit:
*RST<n>