8 hpal voltage calibration – part i of v, Hpal – AMETEK ReFlex User Manual

Page 313

Advertising
background image

Active Loads

ReFlex Power™

7-38

M380056-01 Rev J

7.6.8

HPAL

V

OLTAGE

C

ALIBRATION

P

ART

I

OF

V

This procedure calibrates the voltage for voltage gain 8 and current range
high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.

This procedure requires maximum voltage 54V and maximum current
11Amps.

1. Turn the active load on:

INP<n>:STATE 1

2. Set the conductance count to 0x33333:

CAL<n>:INP:COND:COUNTS 0x33333

3. Set the input DC voltage to 13 Volts. After the input DC voltage

stabilizes enter the first exact voltage:

CAL<n>:INP:VOLT 1 <value-1>

4. Set the input DC voltage to 54 Volts. After the input DC voltage

stabilizes enter the second exact voltage:

CAL<n>:INP:VOLT 2 <value-2>

5. Send the calibration command to calculate the gain and offset:

CAL<n>:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful:

SYST:ERR?

7. If there is no error returned at the system error query command then

save the calibration result as follows:

CAL<n>:UNLOCK “6867”

CAL<n>:STORE

CAL<n>:LOCK

8. Reset the unit

*RST<n>