30 lpal resistance calibration – part iv of v, Lpal – AMETEK ReFlex User Manual
Page 353

Active Loads
ReFlex Power™
7-78
M380056-01 Rev J
7.6.30
LPAL
R
ESISTANCE
C
ALIBRATION
–
P
ART
IV
OF
V
This procedure calibrates the resistance for voltage gain 1 and current
range high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Connect the
sense positive-to-positive lead and sense negative-to-negative lead. Also
connect the shunt and the DC current meter between the active load and
DC power supply
This procedure requires maximum voltage 100.0 V and maximum current
0.5 Amps.
1. Set the DC input voltage to 100 Volts
2. Turn the active load
INP<n>:STATE 1
3. Set the conductance count to 0xC7C.
CAL<n>:INP:COND:COUNTS 0xC7C
4. After DC voltage stabilizes. Enter the voltage using SCPI
commands
CAL<n>:INP:VOLT 1 <value-1>
5. After DC current stabilizes. Enter the current using SCPI commands
CAL<n>:INP:CURR 1 <value-1>
6. Set the conductance count to 0x4EC.
CAL<n>:INP:COND:COUNTS 0x4EC