AMETEK ReFlex User Manual
Page 156

ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-71
14. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:CURR:FIVEPOINT2 <value-2>
15. Program the output current to 50% of the full scale current:
CAL<n>:OUTP:CURR:COUNTS 2276
16. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:CURR:FIVEPOINT3 <value-3>
17. Program the output current to 72.5% of the full scale current:
CAL<n>:OUTP:CURR:COUNTS 3527
18. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:CURR:FIVEPOINT4 <value-4>
19. Program the output current to 95% of the full scale current:
CAL<n>:OUTP:CURR:COUNTS 4451
20. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL<n>:OUTP:CURR:FIVEPOINT5 <value-5>
21. Turn the unit off:
OUTP<n>:STAT OFF
22. Perform the voltage calibration:
CAL<n>:OUTP:CURR:CALC
23. Send the system error query command to make sure calibration was successful:
SYST:ERR?
24. If there is no error returned at the system error query command then save the
calibration results:
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
25. Optional Commands:
Query the
current
calibration points:
CAL<n>:OUTP:CURR:FIVEPOINT ?