23 lpal voltage calibration – part ii of v, Lpal – AMETEK ReFlex User Manual
Page 339

Active Loads
ReFlex Power™
7-64
M380056-01 Rev J
7.6.23
LPAL
V
OLTAGE
C
ALIBRATION
–
P
ART
II
OF
V
This procedure calibrates the voltage for voltage gain 4 and current range
high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.
This procedure requires maximum voltage 109 V and maximum current 3.0
Amps.
1. Turn the active load on
INP<n>:STATE 1
2. Set the conductance count to 0x6666
CAL<n>:INP:COND:COUNTS 0x6666
3. Set the input DC voltage to 27 Volts. After the input DC voltage
stabilizes enter the first exact voltage using following command
CAL<n>:INP:VOLT 1 <value-1>
4. Set the input DC voltage to 109 Volts. After the input DC voltage
stabilizes enter the second exact voltage using following command
CAL<n>:INP:VOLT 2 <value-2>
5. Send the calibration command to calculate the gain and offset:
CAL<n>:MEAS:OFFSET:CALC
6. Send the system error query command to make sure calibration was
successful.
SYST:ERR?